  Aperture Bound tester  KO-300




* Measures a diaphragm operating time and an error of a diaphragm.
Time until it goes into the range of the iris diaphragm specified from the trigger (ms), And the iris diaphragm error (ΔEV) of the specified time (time to be stabilized) is analyzed, A digital display is carried out.
Drive a diaphragm from the outside.(optional)
Measurement with various postures is possible.
Specifications
Measurement Subject Lens containing an iris diaphragm
Measuring item
Refer to measurement chart
T1(ms) Time until bound reaches a fiducial point from an external trigger
T2(ms) Time until bound settles down from an external trigger
ΔEV(FNo. eroor) FNo. error after T3ms (stable point) since external trigger
FNo. indication Indicates standard FNo. of FNo. error(ΔEV)
Has MANUAL and AUTO mode
Measurement chart
Measurement
and
indication range
* Indication method 4 ch digital,7 segment indication
* T1 ch 0.02-99.9ms
* T2 ch 0.02-99.9ms(resolving power 25µs)
* ΔEV ch ±2.99EV(resolving power 0.01EV)
* FNo. ch F1.4--32,&-0.5EV
* T3 setting range 1-99ms,Digital 2fig. indications
* Bound tolerance level 0-+0.99EV,0-0.99EV
Digital 2fig. indications
* FNo. setting range F1.4--32 2times ,Passing
With a switch(-0.5EV shift)
* Measuring aperture range Φ75.0-0.5(m/m)
Mesuring accuracy T1,T2 :±50µs
ΔEV :±0.05EV
Trigger input signal TTL, trigger signal or synchronous(MAKE) signalTTL, trigger signal or synchronous(MAKE) signal
Calibration
method
One calibration of fixed gauge(aperture)
WAVE OUT Wave Form
(ΔEV:0.00EV → about 1.25V,MAX → 12V)
DATA OUT Centronics
Mean, Max, Min (continuous data and five data)
Environment for use Temperature 5ºC-45ºCHumidity below 85%
Power supply AC100V±10%,50/60 Hz
(AC115,220V With a tap)
Power consumption about 50VA
Dimensions&
Weight
[ Main unit ] about 8.0 Kgs
250(W)x210(H)x350(D)mm
[ Detector unit ] about 10.0 Kgs
240(W)x710(H)x200(D)mm